1071Ex
Intrinsically-Safe Ultrasonic Wall Thickness Gauge
With
the 1071-Ex measurements of wall thicknesses can be made quickly and
accurately in the range of 0,7 to 400 mm. There are three different
probe availables to cover this wide measuring range, each of which can
be easily and conveniently changed inside the hazardous area.
- standard probe : 1.5 to 250 mm
- mini probe: 0.7 to 25 mm
- low frequency probe : 0.7 to 25 mm
Due to its compact size and operating time of over 200 hours this easy
to use instrument is essential whenever the continuous supervision of
minimum wall thicknesses is required.
For example, it is possible to check the thickness of protective coats
or for the erosion of pipe or tank walls even while the plant is
operating.
By using the limit value guides, minimum and maximum values can be set and are registered by an audible signal.
The calibration functionality of the instrument means that
adjustmentscan be made to compensate for the type of material being
measured, this is done by one of two ways:
- entering the sound velocity of the corresponding material
- measuring a material with known thickness and adjusting the instrument to that reference
These adjusted parameters then stay in the memory even when switching off the instrument.
FEATURES
- limit value guides and monitoring
- adjustable sensitivity
- easy to use
- energy saving mode
- accuracy : ± 0.05
mm
- Class I Zone 0 AEx ia II C T4 I.S.
Class I Division 1 Group A-D T4
COMPLETE KIT
- Wall thickness measuring gauge 1071-Ex
- Incl. primary cells, case, user manual
- 100 ml of couplant
ACCESSORIES
Optional available :
- standard probe
- mini probe
- low frequency probe
- 500 ml of couplant
- handle-collar
- vulcollan films (10 pcs)
|
| Specifications |
| MODEL |
1071Ex |
| EC - CERTIFICATE OF CONFORMITY |
PTB 97 ATEX 2231 X |
| FM APPROVED |
Class I Zone 0 AEx ia II C T4 I.S. Class I
Division 1 Group A-D T4 |
| PROTECTION |
IP 54 |
| DIMENSIONS |
120 x 65 x 25 mm |
| WEIGHT |
150 g |
| OPERATING TEMPERATURE |
- 10° .....50°C |
| STORAGE TEMPERATURE |
- 20°......60°C
|
| SUPPLY |
2 x AA |
| WORKING TIME |
200 hours continuous operation
1-2 years in stand-by |
| INSTRUMENT |
operation permitted in zone 1 |
| COUPLANT |
perfect connection between sensor and material |
|